Metadata-Version: 2.4
Name: et43
Version: 0.1.0
Summary: Human-friendly Python client for EastTester ET43x LCR meters (ET430-ET433)
Keywords: lcr,lcr-meter,scpi,easttester,et43,et433,et43x,pyserial,instrument
Author: Vyacheslav Mayorov
Author-email: Vyacheslav Mayorov <vyacheslav.mayorov@gmail.com>
License-Expression: MIT
License-File: LICENSE
Classifier: Development Status :: 4 - Beta
Classifier: Intended Audience :: Developers
Classifier: Intended Audience :: Science/Research
Classifier: Operating System :: OS Independent
Classifier: Programming Language :: Python :: 3
Classifier: Programming Language :: Python :: 3 :: Only
Classifier: Programming Language :: Python :: 3.14
Classifier: Topic :: Scientific/Engineering
Classifier: Topic :: System :: Hardware
Classifier: Typing :: Typed
Requires-Dist: pyserial>=3.5
Requires-Python: >=3.14
Project-URL: Homepage, https://github.com/acc15/et43
Project-URL: Repository, https://github.com/acc15/et43
Project-URL: Issues, https://github.com/acc15/et43/issues
Description-Content-Type: text/markdown

# et43

A human-friendly Python client for **EastTester ET43x** handheld LCR meters
(ET430/430B/431/432/**ET433**) over their USB virtual serial port.

> Unofficial, not affiliated with EastTester / Hangzhou Zhongchuang.

Instead of raw SCPI strings you get typed properties and methods: `mode`,
`frequency`, `level_mv`, `equivalent`, `fetch()` and so on.

- Single dependency: [`pyserial`](https://pypi.org/project/pyserial/) (no PyVISA).
- Fully typed (`py.typed`).
- Tested against a real ET433 (firmware `V2.00.2414.001`); see
  [Firmware quirks](#firmware-quirks).

### ET44x / ET45x

The benchtop ET44x/ET45x series speaks a very similar SCPI dialect, so most of
the client may work there too, but it has not been tested. Known differences,
based on the vendor's ET44/ET45 SCPI manual and the
[ET44.py](https://github.com/philpagel/ET44.py) driver:

- auto range is `FUNC:IMP:RANGe:AUTO ON|OFF` there, while the ET433 needs
  `FUNC:IMP:RANGE AUTO`, so `measurement_range` will likely fail;
- ET44.py sends `THETA` for phase, which the ET433 rejects (it needs `THR`);
- the default baud rate is 9600 (`discover()` tries it as well).

Reports from ET44x/ET45x owners are welcome.

## Installation

```bash
pip install et43
# or
uv add et43
```

Requires Python 3.14+.

## Quick start

```python
from et43 import ET43, Equivalent, PrimaryMode, SecondaryMode

with ET43.open("/dev/ttyUSB0").auto_remote() as lcr:
    print(lcr.idn)

    # Capacitance and ESR at 1 kHz, 1 V test level, series model
    lcr.mode = PrimaryMode.C
    lcr.secondary = SecondaryMode.ESR
    lcr.frequency = 1_000
    lcr.level_mv = 1_000
    lcr.equivalent = Equivalent.SER

    primary, secondary = lcr.fetch()
    print(primary, secondary)  # 1.234e-06 0.5
```

Without a port name the client looks for the instrument itself:

```python
lcr = ET43.open()  # auto-discovery, tries 115200 and 9600 baud
```

`ET43.open()` is a convenience factory: it creates and configures a
`serial.Serial` and hands it to the client. If you already have a port object,
pass it directly:

```python
import serial
from et43 import ET43

link = serial.Serial("/dev/ttyUSB0", 115200, timeout=2)

with ET43(link).auto_remote() as lcr:
    primary, secondary = lcr.fetch()
```

The client always owns the port it is given: `close()` and leaving a `with`
block close it, even if you created the `serial.Serial` yourself. The response
timeout is taken from `link.timeout` (2 seconds if it is `None`). The port
object is available as `lcr.port`, and the device name as `lcr.port.port`.

The `ET43(link)` constructor performs no I/O. The first exchange happens on
`remote()`, `fetch()` or another explicit call.

On POSIX the port is locked with `flock` by default, so a second process gets
`serial.SerialException` instead of corrupting the exchange. Disable this with
`ET43.open(..., exclusive=False)`.

## Settings

Every setting can be read and written:

```python
from et43 import DisplayPage, Equivalent, PrimaryMode, SecondaryMode, Speed

lcr.mode = PrimaryMode.C
lcr.secondary = SecondaryMode.ESR
lcr.frequency = 10_000  # Hz
lcr.level_mv = 600  # mV
lcr.bias_mv = 100  # mV
lcr.equivalent = Equivalent.SER
lcr.speed = Speed.MED
lcr.measurement_range = None  # None = AUTO, a number = range in ohms
lcr.relative_mode = True
lcr.display_page = DisplayPage.SYST
```

Enum properties (`mode`, `secondary`, `equivalent`, `speed`, `display_page`)
accept enum members only, and their getters return enum members. They are
`StrEnum`s, so comparing with a string still works: `lcr.mode == "C"`.

`measurement_range` reads as `int | None`: `None` means `AUTO`, otherwise it is
the range resistance in ohms.

`frequency`, `level_mv` and `bias_mv` read as `float`, because SCPI returns them
in NR3 format. In practice the ET433 works in 1 Hz / 1 mV steps.

The client does not validate ranges: values are sent to the instrument as is.
Valid ranges depend on the model, and the instrument decides how to handle an
invalid value. For example, the ET433 silently ignores an out-of-range frequency
or bias, and answers `exec err` to an invalid level or measurement range.

Parsing helpers on the enums:

```python
DisplayPage.SYST  # "SYST", the SCPI token; the instrument answers "SYSTem"
PrimaryMode.from_string("c")  # PrimaryMode.C
PrimaryMode.from_string("Cp")  # PrimaryMode.C (older firmware answers)
Speed.from_string("MEDium")  # Speed.MED (instrument answer)
```

## Measurements

`fetch()` sends exactly one `FETCH?` query and returns a `Reading`, a
two-field `NamedTuple`. Unpack it or use the field names:

```python
primary, secondary = lcr.fetch()

reading = lcr.fetch()
print(reading.primary, reading.secondary)
```

- `primary` is always a `float`.
- `secondary` is a `float` or `None`: in `DCR` mode, for example, the
  instrument returns a single value. The tuple always has two items, so
  `primary, secondary = ...` works in every mode.

`FETCH?` does not report the mode or units, so the client makes no extra
queries. Set the mode, the equivalent circuit and other parameters through
properties:

```python
import time

lcr.mode = PrimaryMode.C
lcr.secondary = SecondaryMode.ESR
lcr.frequency = 1_000

time.sleep(0.5)  # let the instrument settle
primary, secondary = lcr.fetch()
```

`fetch()` neither configures the instrument nor waits: it is a single query.
Measurement loops and frequency sweeps are left to your code; see
[`scripts/`](https://github.com/acc15/et43/tree/master/scripts) in the repository for examples.

## Remote and local mode

Neither `ET43.open()` nor `ET43(link)` switches the instrument to remote.
There are two ways to use the client:

- a plain `with` block: the instrument stays in local mode;
- `auto_remote()`: remote mode for the duration of the block, a guaranteed
  return to local mode however the block ends, then the port is closed.

```python
with ET43.open("/dev/ttyUSB0") as lcr:
    print(lcr.fetch())  # local mode

with ET43.open("/dev/ttyUSB0").auto_remote() as lcr:
    print(lcr.fetch())  # remote mode; SYST:LOCAL and close() afterwards
```

`auto_remote()` sends `SYST:REMOTE` on entry and `SYST:LOCAL` on exit, also when
the block raises. This matters: in remote mode the front panel is locked, and it
must not stay locked after your program ends. If an error occurs while
returning to local mode after another error, the original error is the one you
see. Call it right on `open()`, since it closes the port on exit. To keep the
connection open across blocks, call `remote()` and `local()` yourself.

In local mode the instrument answers queries (`FETCH?`, `FREQ?` and so on) but
**ignores configuration commands**. The client reports this as an error rather
than pretending the setting was applied:

```python
with ET43.open("/dev/ttyUSB0") as lcr:
    print(lcr.fetch())  # queries work
    lcr.frequency = 10_000  # raises TimeoutError in local mode
```

## Calibration

```python
lcr.calibrate()  # CORR:EXEC
```

The instrument detects what is connected, open or short, by itself. Connect open
leads and calibrate, then short the leads and calibrate again.
**This overwrites the stored open/short correction data.**

`calibrate()` waits up to 45 seconds for the `exec success` acknowledgement
(a real ET433 takes about 45 seconds). The timeout can be changed:

```python
lcr.calibrate(timeout=60)
```

## Thread safety

The client is not thread-safe: use one instance from one thread. If several
threads need the instrument, serialize calls with your own `threading.Lock`,
for example around `fetch()` and property writes.

## Firmware quirks

Verified on a real ET433 with firmware `V2.00.2414.001`:

- **Frequency:** 10–100000 Hz. 10 Hz really works and gives correct C/ESR on a
  2.2 µF electrolytic capacitor. Out-of-range values are ignored: the instrument
  answers `exec success` but keeps the old frequency. The ET432 range differs.
- **Bias:** 0–1000 mV. The older PDF says 0–500 mV, but the V2 manual and the
  instrument allow 0–1000 mV. Out-of-range values are ignored, fractional
  values are truncated to whole millivolts.
- **ECAP mode** enables an internal 1 V bias, which may stay on after switching
  to another mode. Check `lcr.bias_mv` if it matters.
- **Level:** 0–1100 mV in whole millivolts. `1101`, negative and fractional
  values (including `0.001` mV) are rejected with `exec err`.
- **Baud rate:** 115200. The manual mentions 9600, so `discover()` tries both.
- **`FETCH?`** returns **two** comma-separated values (`primary,secondary`),
  while the older SCPI PDF describes four. In `DCR` mode it returns **one**
  value, and `Reading.secondary` is `None`. Mode and units are not included.
- **Line terminators** vary between responses (`\n` and `\r`); the reader
  accepts both.
- **Parallel equivalent circuit** is spelled `PAL`: `FUNC:IMP:EQU PAR` is
  rejected, and `FUNC:IMP:EQU?` answers `SERial` or `PALlel`. This is an
  EastTester misspelling of "parallel", shared by the ET44x/ET45x and present
  in their SCPI manual. The API member is `Equivalent.PAR`, and its value is
  `"PAL"`, the token the instrument accepts. `from_string` understands `PAR`,
  `PAL` and `PALlel`.
- **Phase:** `FUNC:IMP:B THETA` is rejected; use `THR`. Phase is returned in
  **degrees**.
- **Auto range:** `FUNC:IMP:RANGE:AUTO` does not work; the correct form is
  `FUNC:IMP:RANGE AUTO`.
- **`AUTO` mode** does not report which parameter was actually measured, and
  `fetch()` returns bare numbers. Set the mode explicitly (`C`, `L`, `R`, `Z`)
  if you need units and meaningful labels.

## Development

The project uses [uv](https://docs.astral.sh/uv/) and
[just](https://github.com/casey/just).

```bash
uv sync       # create .venv with the package and dev tools
just check    # ty + ruff + pytest
just build    # check, then build sdist and wheel into dist/
```

The tests do not need an instrument. The example scripts do:

```bash
uv run scripts/basic.py
uv run scripts/sweep_csv.py
```

Project layout:

```
src/et43/
  client.py     # ET43: transport, settings, fetch, remote/local
  enums.py      # PrimaryMode, SecondaryMode, Speed, Equivalent, DisplayPage
scripts/        # usage examples (not part of the package)
tests/
```

## License

[MIT](https://github.com/acc15/et43/blob/master/LICENSE)